Load Cell Testing Procedure: Step-by-Step with Mounting Fixture Setup

As a professional load cell manufacturer, Haibo provides this complete load cell testing procedure to demonstrate how to verify sensor performance using a mounting fixture, load cell indicator, and standard test weights. It covers essential steps from four-wire connections to gradual load application, focusing on the two most critical parameters: sensitivity (mV/V) and zero balance.

Based on our standard factory acceptance criteria, this test is performed on a load cell with a rated sensitivity of 2.4 ±0.5 mV/V. The demonstrated results—2.2491 mV/V for sensitivity and 0.0259 mV/V (approx. 300 counts) for zero balance—confirm that the unit meets our strict quality standards. This guide serves as a practical reference for technicians and engineers to ensure accurate, repeatable testing.

称重传感器测试

Load Cell Working Principle

How does a load cell convert weight into an electrical signal? The core lies in strain gauges and the Wheatstone bridge.

When the sensor is loaded (e.g., weights are applied), the internal metal spring element deforms slightly. Strain gauges bonded to the element stretch or compress, changing their electrical resistance. The four strain gauges form a Wheatstone bridge, which converts these resistance changes into a weak voltage signal (in the mV range).

The weighing indicator receives this voltage signal and, combined with the sensor’s sensitivity (mV/V) and excitation voltage, accurately calculates and displays the actual weight. Testing sensitivity and zero output essentially verifies whether this conversion process is normal and accurate.

称重传感器测试

Equipment and Tools for Load Cell Testing

Preparing the necessary equipment and tools is essential for obtaining reliable measurement data. A typical load cell testing setup consists of a testing fixture for securing and positioning the load cell, a load cell indicator for supplying excitation voltage and measuring output signals, and standard weights for applying known loads. The following is a list of the key equipment and tools required for this test:

称重传感器测试

Testing Platform and Load Cell Testing Fixture

The testing platform provides a stable base for load cell testing, minimizing the effects of external vibration. The load cell testing fixture secures and positions the sensor to ensure proper alignment and consistent loading conditions.

称重传感器测试

Load Cell Indicator - Excitation and Signal Measurement

A load cell indicator supplies excitation voltage and measures the sensor’s output signal. Some models also display output in mV/V, allowing technicians to monitor signal changes under different loads.

称重传感器测试

Terminal Block and Load Cell Cables - Electrical Connections

The terminal block connects the load cell wiring to the testing instrument, ensuring proper electrical connections. For a typical four-wire load cell, the red, black, green, and white wires correspond to Exc+, Exc−, Sig+, and Sig−, respectively. Always follow the manufacturer’s wiring diagram to prevent incorrect connections or short circuits.

称重传感器测试

Standard Weights - Load Application

Standard test weights apply known loads to evaluate the load cell’s output response at different load levels. Weights should be applied gradually and gently to prevent impact loading from affecting test results or damaging the sensor.

Step-by-Step Load Cell Testing Procedure

This step-by-step guide demonstrates how to test a load cell using a load cell indicator, a testing fixture, and standard weights. The procedure covers equipment setup, sensor wiring, mounting, and gradual loading up to the rated capacity. By measuring the output signal at full load and under no-load conditions, you can check the load cell’s sensitivity and zero output against the specified requirements.

Step 1: Power on the indicator

Plug the power cord into the power inlet of the indicator; connect the terminal board to the sensor port at the rear of the indicator.

Step 2: Confirm the fixture

The fixture is pre-mounted in position; no movement or adjustment is required.

Step 3: Wire the sensor

Connect the sensor cable to the terminal board in the sequence red, black, green and white.

Step 4: Mount the sensor

Place the sensor into the mould and tighten the screws so that the sensor is firmly clamped.

Step 5: Load weights and test

Load the weights step by step up to the rated capacity, read the indicator output and confirm that it falls within the acceptable range.

Step 6: Sensitivity & Zero Output

After loading to the rated capacity, read the full-scale sensitivity output (rated 2.4 ±0.5 mV/V); then read the zero output with no load (requirement: within ±0.5 mV/V, counts ≤4999).

Key Parameters in Load Cell Testing: Sensitivity and Zero Balance

Load cell testing involves several key performance parameters used to evaluate measurement accuracy, stability, and reliability. These include sensitivity, zero balance, nonlinearity, hysteresis, and repeatability. In this test, we focus on demonstrating how to measure load cell sensitivity and zero balance. These two parameters help assess the electrical output characteristics of a load cell under unloaded and loaded conditions and identify potential output deviations.

称重传感器测试

Load Cell Sensitivity Test

Load cell sensitivity refers to the output signal generated at rated load relative to the excitation voltage, typically expressed in mV/V. During testing, a rated load is applied, and the instrument reading is recorded to verify whether the measured sensitivity meets the acceptance criteria. In this test, the load cell has a rated sensitivity of 2.4 mV/V, with an internal factory acceptance tolerance of ±0.5 mV/V. The measured sensitivity is 2.2491 mV/V, meeting the acceptance criteria.

称重传感器测试

Load Cell Zero Balance Test

Zero balance refers to the initial electrical output of a load cell under no-load conditions, typically expressed in mV/V. During testing, the load cell is kept unloaded, and the instrument reading is recorded to check whether its initial output is within the acceptable range. For this test, the internal acceptance criteria are ±0.5 mV/V, with an instrument count limit of 4,999. The measured zero balance is 0.0259 mV/V (approximately 300 counts), meeting the acceptance criteria.

Load Cell Testing Video Demonstration

In addition to our step-by-step load cell testing instructions, we provide a complete load cell testing video to demonstrate the actual testing process. From mounting the load cell on the testing fixture and connecting the wiring to applying standard weights and measuring the output signal, the video walks you through each stage of the load cell testing procedure.

By following this practical demonstration, you can better understand how to use load cell testing equipment, read load cell output signals, and perform key performance tests, including zero balance and sensitivity testing. It also highlights important operating details to help you follow the testing process more clearly and accurately.

load cell testing video demonstration

Precautions for Load Cell Testing

Following the correct load cell testing procedure is essential for obtaining reliable and accurate measurement results. Proper wiring, sensor installation, gradual load application, and accurate data recording all play an important role in the testing process. Keep the following precautions in mind to avoid common testing errors and ensure consistent results.

称重传感器测试

1. Check Power Supply and Load Cell Wiring Connections

Before powering on the indicator, verify that the supply voltage matches its rated input voltage. Always disconnect the power before making or changing electrical connections. Make sure the red, black, green, and white wires are connected to their corresponding terminals according to the wiring diagram to prevent incorrect connections or short circuits.

2. Ensure Proper Load Cell Installation

After placing the sensor in the fixture, make sure it is fully seated against the locating surface. Tighten the mounting screws evenly to prevent misalignment or uneven loading.

称重传感器测试
称重传感器测试

3. Apply Standard Weights Gradually

Place standard weights gently and apply them step by step to avoid impact loading. Remove the applied weights after testing.

4. Record and Review Load Cell Output Readings

Record the output readings at each load point for future reference. If any reading falls outside the specified tolerance, stop the test and investigate the cause before proceeding.

称重传感器测试

Load Cell Testing and Quality Control at Haibosensor

Accurate and stable output readings are important indicators of load cell performance. As a direct load cell manufacturer, Haibosensor follows standardized production and quality control procedures, including factory testing of key output parameters, to ensure that our products meet the required quality standards. In this load cell testing demonstration, the measured sensitivity was 2.2491 mV/V, and the zero output also met the specified factory requirements. These results confirm that the tested load cell meets the relevant requirements for these key output parameters. If you have any questions about load cell testing, technical specifications, or product selection, feel free to contact Haibosensor. Our team is ready to provide professional technical support.